Quality Control and Calibration 5-13
14. The analyzer then calculates and displays the new Calibration factors. Select
Accept
to com-
plete the calibration.
15. Select
Exit
.
16. Perform a Quality Control run two or three times to verify that calibration was successful: see
“Running a Quality Control Sample” on page 5-8
.
5.2.4 View Calibration History
All calibration times and factors are saved, and can be viewed as follows:
1. From the Home screen, select Maintenance > Calibration > His-
tory
.
2. Use the on-screen arrows to move through the history.
5.2.5 Resetting Calibration
If calibration fails repeatedly, call Abaxis Technical Support. If directed, use the Reset Calibration
function as follows.
1. From the Home screen, select Maintenance > Calibration > Reset Calibration.
2. Select
OK
.
3. Calibrate the analyzer: see
“Calibration Procedure” on page 5-11
.
Note: If the analyzer displays the message 1 or more of
your factors are out of range
, contact Abaxis
Technical Support.