EasyManua.ls Logo

Abaxis VetScan HM5 - Linearity Ranges

Abaxis VetScan HM5
164 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications 10-3
10.2 Linearity Ranges
The VetScan HM5 is guaranteed to provide specified accuracies within its linearity range when prop-
erly calibrated and maintained. Beyond this range, results may still be displayed, but accuracy is no
longer guaranteed.
If the value is over the maximum range of guaranteed linearity, the instrument cannot measure it, and
the result will be marked with an
E
,
m
,
M
, or
N
flag.
To measure a sample whose parameters exceed the maximum linear value indicated in the table below,
predilution is recommended.
The following tables list the linearity ranges for primary parameters in normal measuring mode and
prediluted mode.
Table 10-1: Linearity Ranges in Normal Measuring Mode
Table 10-2: Linearity Ranges for Prediluted Mode
Parameter Linearity Ranges Maximum Unit
WBC 0–100 150 10
9
cells/liter
EOS 0.2–3.0 10.0 10
9
cells/liter
RBC 0–15 20 10
12
cells/liter
PLT 0–700 1000 10
9
cells/liter
HGB 0–250 400 g/l
HCT 0–100 %
MCV 30–150 Fl
MPV 3–30 Fl
Parameter Linearity Ranges Maximum Unit
WBC 2–200 300 10
9
cells/liter
EOS 0.2–3.0 10.0 10
9
cells/liter
RBC 1–30 40 10
12
cells/liter
PLT 100–2000 3000 10
9
cells/liter

Table of Contents

Other manuals for Abaxis VetScan HM5

Related product manuals