3-10 Settings
3. Set PC Automatic Send to
Enabled
.
4. Select
Accept
.
3.6.4.2 Electronic Medical Record (EMR)
The Host Network > Electronic Medical Record (EMR) settings enable the HM5 to exchange work
order and results (bidirectional) with a network-enabled PC or software such as a practice
management system. Orders flow from the EMR host to the HM5, and results from the HM5 to the
EMR host. The EMR host system software must be capable of bidirectional communication with the
HM5.
When receiving orders, the HM5 establishes a listening service and waits for connections from the
EMR host. The listening address is the HM5 analyzer IP address (
see
“Device Network” on page 3-8),
and the listening port is determined by the EMR Local Port setting (see below).
When sending results, the HM5 establishes an outbound connection to the specified IP address and
port (see below), and the receiving EMR system must be running a service listening for incoming
connections on that IP address and port. The IP address and port values are provided by the local
network administrator or software vendor.
To configure the HM5 to communicate with an EMR system:
1. In the Host Network screen, set Electronic Medical Record (EMR) to Active.
2. Enter the EMR Remote IP Address and EMR Remote Port number, as provided by your local
network administrator or software vendor.
3. Enter the EMR Local Port number provided by, or determined in conjunction with, the EMR
software vendor. (This is the port the HM5 listens on for incoming order connections.)
4. Set EMR Automatic Send to
Enabled
.
5. Press
Accept
.
Note: Enabling PC Automatic Send is typical for this configuration,
but may not be appropriate for your PMS or individual configu-
ration. If you have questions about the correct setting for your
clinic, contact Abaxis Technical Support or consult your PMS
vendor.
Note: Enabling EMR Automatic Send is typical for this configuration,
but may not be appropriate for your PMS or individual configu-
ration. If you have questions about the correct setting for your
clinic, contact Abaxis Technical Support or consult your PMS
vendor.