EasyManua.ls Logo

ABB AW101

ABB AW101
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Water Hardness Analyzer
AW101 Testomat 4 Description of Displays and Operating Features
6 IM/AW101 Rev. C
4 Description of Displays and
Operating Features
a Status of Limit Value Displays
Displays the status of the limit values LV1 and LV2.
b Text Display
Displays the current analysis, all important status results
and programming data in a 2-line LCD.
c Alarm
Displays a function fault.
d Power Switch
The On/Off switch is located on the right-hand side panel.
e Unit Fuse (inside the unit)
Protects outputs against overload and short circuit.
f Analysis Message
Displays current analysis.
g Programming Keys (Cursor block with ENTER)
These keys are used to enter all values and programming
data.
h 'Manual' – manual start of an analysis.
i 'STANDBY' – manual analysis stop/standby.
j 'Alarm' – cancels alarm message.
k I-Key
Access all unit information.
l M-key
Access the programming menu.
4.1 Display Functions
Status of limit values 1 and 2
1: The red display lamp is lit if limit value 1 i(LV1) s
reached or exceeded. It turns green if the measured value
falls below the set limit value.
2: The red display lamp is lit if limit value 2 (LV2) is
reached or exceeded. It turns green if the measured value
falls below the set limit value.
Status and measured value display:
In continuous operation the current measured value is
shown in the upper line.
When the measured value falls below the measuring range
’<’ is displayed: for example, <0.05 °dH
When the measured value exceeds the measuring range
’>’ is displayed: for example, >10.0 °dH
When the analysis interval is stopped while in operation
(Analysis Stop), STANDBY and measured value are
displayed alternately.
The icon appears if the low-level reagent alarm is
active.
4.2 Limit Value Displays
The adjustable limit values are shown in the bottom display line.
4.3 Alarm Message
A fault condition is signalled by a red light.
Error messages are displayed alternately with the normal display
text and can only be deleted by cancellation and correction of
the fault.
Fig. 4.1 Display Functions
0.12
^ 1:0,30 ^ 2:0,40
0.12
^ 1:0,30 ^ 2:0,40
ANALYSE
1
2
ANALYSE
°d
H
1
2
°d H
5
4
1
2 3
12
9
11
10
8
7
6
M
I
Status of
limit value
Permanent
measurement
display
Adjustable
limit values
Symbol:
Reagent
alarm: ON
Indicators for
alarm and
maintenance

Related product manuals