16 EL3000 CONTINUOUS GAS ANALYZERS | CI/EL3000-EN REV. B
Requirements for the installation site, power supply
NOTE
For the analyzers ZO23 and Fidas24 the information contained in the sec-
tions "ZO23: Preparation for installation" (see page 28) or "Fidas24: Prepara-
tion for installation" (see page 34) must be considered in addition.
Requirements for the installation site
The gas analyzer is only intended for installation indoors.
The technical data of the gas analyzer (see data sheet) are valid for an in-
stallation location altitude up to 2000 meters above sea level.
The installation site must be stable enough to bear the weight of the gas
analyzer! To ensure safe installation and removal, we recommend that the
19-inch housing is supported in a cabinet or rack with slide rails!
Short gas paths
Install the gas analyzer as close as possible to the sampling location.
Install the gas conditioning and calibration modules as close as possible to
the gas analyzer.
Adequate air circulation
Provide for adequate natural air circulation around the gas analyzer. Avoid
heat build-up.
Mount (see page 81) several 19-inch housings with a minimum spacing of
1 height unit between housings.
Protection from adverse conditions
Protect the gas analyzer from:
Cold
Exposure to heat from e.g. the sun, furnaces, boilers
Temperature variations
Strong air currents
Accumulation of dust and ingress of dust
Corrosive atmospheres
Vibrations
Climatic conditions
Relative humidity max. 75 %, no condensation
Ambient temperature
for storage and transport −25 to +65 °C
during operation +5 to +45 °C
Uras26 in combination with another analyzer,
Limas23, Fidas24
+5 to +40 °C