96 EL3000 CONTINUOUS GAS ANALYZERS | CI/EL3000-EN REV. B
Electrical connections Analog output modules
Pin Signal
1 AO1+
2 AO1
3 AO2+
4 AO2
5 AO3+
6 AO3
7 AO4+
8 AO4
Analog outputs AO1 to AO4
0/4 to 20 mA (pre-set ex works to 4 to 20 mA), common negative pole, elec-
trically isolated to ground, can be connected to ground as required, in this
regard, max. gain compared to local protective ground potential 50 V, work-
ing resistance max. 750 Ω. Resolution 16 bit. The output signal may not be
less than 0 mA.
Design
4-pole or 8-pole plug-in terminal strip. Please refer to the information about
the requisite material (see page 101).
Terminal layout
An analog output is allocated in the sequence of the sample components for
each sample component. The sequence of the sample components is docu-
mented in the analyzer data sheet and on the type plate.
NOTE
The assignment of the terminals can be changed in the configurator.