Endura AZ series integral and remote transmitter
Combustion oxygen monitor 5 Programming
28 IM/AZ20E–EN Rev. F
5.2.2 Operator Menu
The Operator Menu is used to view:
a list of current alarms (from the Diagnostics option)
Operator Page 1 only, Operator Page 2 only or, if
Autoscroll is enabled, the two page views are toggled at
10-second intervals.
active signals and their values (from the Signals View
option)
Note. It is not possible to configure the system or to modify
data from the Operator Menu.
Display Overview Section and Function
Diagnostics
From the Operator Menu, press to select the Diagnostic
menu that displays active diagnostic system data.
Use the and keys to scroll through
active diagnostics.
Refer to Section 8.2.1, page 79 for diagnostic messages.
Operator Page 1 (2)
Tw o Operator pages are available – Operator Page 1 is always
enabled, Operator Page 2 can be set to OFF or enabled.
When Operator Page 2 is enabled, it is listed in the
Operator Menu.
Setup of the two Operator pages is performed at the Display
menu by selecting pre-configured display options – see
Section 5.4.5, page 48.
Autoscroll
When enabled the 2 Operator pages (1 and 2) are toggled at
10-second intervals, providing on-screen monitoring of up to 6
parameters.
Once selected, the transmitter remains in Autoscroll mode
until Operator Page 1 or Operator Page 2 is selected. If
Operator Page 2 is set to Off, or AutoScroll is set to Disabled,
Operator Page 1 is displayed.
Table 5.3 Operator Menu Screens
1
Operator Menu
Back
Select
Diagnostics
Operator Page 1
Operator Page 2
Autoscroll
Signals View
Sensor
–– F098.001 ––
TEMP SENSOR
FAILED
Check related
diagnostics
check wiring.
Back
Exit
Sensor
O2
TMP
O/P
%O2
°C
0%
21.0
12.00
4
Operator Menu
Select
Diagnostics
Operator Page 1
Operator Page 2
Autoscroll
Signals View

Table of Contents