Endura AZ series integral and remote transmitter
Combustion oxygen monitor 5 Programming
36 IM/AZ20E–EN Rev. F
5.4.2 Calibrate
Used to calibrate the sensor, select test gas types, set oxygen
hold action, enable automatic calibration hardware (if fitted)
and set calibration diagnostics options.
Note. Ensure reference air and test gas supplies are set up
before running a calibration – see IM/AZ20P–EN.
Parameter Comment / [Range] Default
Sensor Cal. See Section 6.1, page 64 for calibration routines.
Note. Calibration options 1-Pt AutoCal to Manual Cal -
2-Pt are displayed only when the probe reaches 690 °C
(1274 °F) and all error messages are cleared.
Restore Defaults is displayed only when the probe
temperature is < 690 °C (1274 °F).
N / A
1-Pt AutoCal Performs a single-point automatic calibration.
Enabled only if calibration is currently inactive, Autocal is
fitted and temperature has stabilized.
2-Pt AutoCal Performs a two-point automatic calibration.
Enabled only if calibration is currently inactive, Autocal is
fitted and temperature has stabilized.
Manual Cal – Offset Performs a manual single-point offset calibration.
Enabled only if calibration is currently inactive and when
temperature has stabilized.
Manual Cal – Factor Performs a manual single-point factor calibration.
Enabled only if calibration is currently inactive and when
temperature has stabilized.
Manual Cal – 2-Pt Performs a manual two-point calibration.
Enabled only if calibration is currently inactive and when
temperature has stabilized.
Restore Defaults Restores the factory calibration offset and factor values
entered in Easy Setup – see page 35.
Table 5.7 Calibrate Menu