Endura AZ series integral and remote transmitter
Combustion oxygen monitor 5 Programming
IM/AZ20E–EN Rev. F 61
5.4.8 Communication
Used to configure HART communication parameters and
cyclic output communication via the IrDA interface on the front
of the transmitter.
Parameter Comment / [Range] Default
HART
HART Poll Address Used to enter a unique device address.
When set to 0, Current Output 1 is active and provides
an analog output proportional to its source. HART
operates in point-to-point mode.
If set between 1 and 15, Current Output 1 is fixed at 4
mA and HART operates in multi-drop mode.
Note. See Section 7.2, page 69 for multi-drop
configuration details.
[0 to 15]
0
HART Device ID The unique ID for the HART device. Factory-set
read-only ID
HART Tag Used to enter a user-defined tag for the HART device
(transmitter), selected from the alphanumeric character
set displayed when the Edit button is selected at this
sub-parameter.
[0 to 8 characters]
N / A
HART Device Message Used to specify a a user-defined message for the HART
device (transmitter), selected from the character set
displayed when the Edit button is selected at this
sub-parameter.
[0 to 16 characters]
N / A
HART Descriptor Used to specify a unique HART descriptor, selected from
the character set displayed when the Edit button is
selected at this sub-parameter.
Maximum of 32 characters.
N / A
Device Install Date The time and date of installation. User-set
Final Assembly No. The device's (transmitter’s) final assembly (ID) number. Factory-set
Table 5.13 Communication Menu