Endura AZ series integral and remote transmitter
Combustion oxygen monitor 5 Programming
34 IM/AZ20E–EN Rev. F
…Easy Setup
Cable Length The length in metres of the cable between the probe and
the remote transmitter. This length is used in the cold
junction measurement to compensate for the impedance of
the cable.
[0 to 100 m]
Note. Applicable only to remote transmitters.
0 m
Autocal Hardware Selects the type of automatic calibration hardware to be
used.
None
None All auto-calibration functions disabled.
Internal Select this option if the probe's (optional) built-in automatic
calibration is fitted.
Remote Select this option if an external automatic calibration system
is to be used.
Note. The solenoid valve outputs or relays can be used to
activate a remote automatic calibration system.
%O2 Range Hi Sets the maximum oxygen concentration.
[0.01 to 100 % O
2]
25.00 %O
2
%O2 Range Lo Sets the minimum oxygen concentration.
[0.01 to 100 % O
2]
0.01 %O
2
Thermocouple Type Selects the thermocouple type used for cell temperature
measurement.
For heated probes (AZ20 and AZ30) this is fixed as Type K.
For unheated probes (AZ25 and AZ35), select the
thermocouple type used in the probe being used with the
transmitter.
K [–100 to 1300 °C (–148 to 2372 °F)] K
B [–18 to 1800 °C (–0.4 to 3272 °F)]
N [–200 to 1300 °C (–328 to 2372 °F)]
R [–18 to 1700 °C (–0.4 to 3092 °F)]
S [–18 to 1700 °C (–0.4 to 3092 °F)]
Parameter Comment / [Range] Default
Table 5.6 Easy Setup Menu (Continued)