Endura AZ series integral and remote transmitter
Combustion oxygen monitor 6 Calibration
IM/AZ20E–EN Rev. F 67
…Calibration Procedures
New calibration
coefficients
discarded
Accuracy
check entry added
to Performance
Log
•Return to Sensor
Cal.
• New calibration
coefficients saved
Save successful
Calibration
entry added to
Performance Log
• Return to Sensor
Cal.
Save unsuccessful
Accuracy
check entry added to
Performance Log
• Display Failure
Calibration completed successfully
1-Pt AutoCal
2-Pt AutoCal
Manual Cal – Offset
Manual Cal – Factor
Manual Cal – 2-Pt
•Return to Sensor
Cal.
Calibration failed to complete
The reason for failure is displayed:
– stability criteria
not met
– no response to the application
of a test gas detected
– calibration successful but Save of
new calibration factor unsuccessful
– newly
calculated calibration factor is outside the limits defined
by Factor Limit and / or the newly calculated calibration
offset is outside the Offset Limit parameter – see page
42.
1-Pt AutoCal
2-Pt AutoCal
Manual Cal – Offset
Manual Cal – Factor
Manual Cal – 2-Pt
<Cal Type>
O2 1.00 %O2
Cal Offset 0.00 mV
Cal Factor 1.000
Reject
Accept
<Cal Type>
O2 20.95 %O2
Exit
Calibration Failed
Signal Not Stable
Signal Not Stable
No Response
NV Error
Result Out of Bounds