FEM630 | ELECTROMAGNETIC FLOWMETER  |  OI/FEM630-EN REV. A
6     ...Electrical connections
1  Use the supplied signal cable to connect the sensor and 
the transmitter.
2  Connect the signal cable in the terminal box of the 
sensor. 
3  Route the cable from the terminal box to above the 
maximum flooding height of 16.4 ft (5 m).
4  Tighten the cable gland.
5  Carefully seal the terminal box. Make sure the gasket for 
the cover is seated properly. 
Note
As an option, the sensor can be ordered with the signal cable 
already connected to the sensor and the terminal box already 
potted.
Potting the terminal box on-site
If the terminal box is to be potted subsequently on-site, a 
special two-component potting compound can be ordered 
separately (order no. D141B038U01). Potting is only possible if 
the sensor is installed horizontally. Observe the following 
instructions during work activity:
Preparation
•  Complete the installation before potting in order to avoid 
moisture penetration. Before starting, check all the 
connections for correct fitting and stability
•  
compound away from the O-ring and the gasket / groove 
(see Figure 37).
•  Prevent the two-component potting compound from 
penetrating the cable conduit (Conduit) for an ½ in NPT 
installation (if used).
Procedure
...Connection with IP rating IP 68
Electrical connection
NOTICE
Adverse effect on the IP rating IP 68
The IP rating IP 68 of the sensor may be adversely 
affected as a result of damage to the signal cable.
•  The sheathing of the signal cable must not be 
damaged.
Danger to health!
The two-component potting compound is toxic – observe 
all relevant safety measures!
Comply with the safety data sheet of the two-component 
potting compound before preparations are started.
Risk notes:
•  R20: Damaging to health when inhaled.
•  R36/37/38: Irritates the eyes, respiratory organs and 
the skin.
•  R42/43: Sensitization through inhaling and skin 
contact is possible.
Safety advice:
•  S23: Do not inhale gas/smoke/humidity/aerosol.
•  S24: Avoid contact with the skin.
•  S37: Wear suited protective gloves.
•  S63: In case of an accident due to inhaling: take the 
injured person out into the fresh air to rest.
CAUT ION
1  Cut open the protective enclosure of the two-component 
potting compound (see packing).
2  Remove the connection clamp of the potting compound.
3  
reached.
4  Cut open the bag at a corner. Perform work activity 
within 30 minutes.
5  Carefully fill the terminal box with the two-component 
potting compound until the connection cable is covered.
6  Wait a few hours before closing the cover in order to 
allow the compound to dry, and to release any possible 
gas.
7  Ensure that the packaging material and the drying bag 
are disposed of in an environmentally sound manner.
G12152
1
2
3
4
5
Packaging bag
Connection clamp
Two-component potting compound
Drying bag
Maximum fill level
1
2
3
4
5
Figure 37  Terminal box sealing