37
FEM630 | ELECTROMAGNETIC FLOWMETER  |  OI/FEM630-EN REV. A
6     ...Electrical connections
...Pin assignment
Electrical data for inputs and outputs
Note
•  An additional document with Ex safety instructions is 
available for measuring systems that are used in 
potentially explosive atmospheres.
•  Ex safety instructions are an integral part of this manual. 
As a result, it is crucial that the installation guidelines and 
connection values it lists are also observed. The icon on 
the name plate indicates the following: 
Current output Uco / 32, 31 / 32
Can be configured for outputting mass flow and volume 
flow via the on-site software.
Power supply
*Source voltage U depends on the load R
B
 and must be within the 
permissible range.
For information on communication via the HART protocol, 
refer to HART® communication on page 48.
AC power supply
Terminals
L / N
Operating voltage
Power consumption
S
Power-up current
DC voltage supply
Terminals
Operating voltage
Ripple
Power consumption
Power-up current
Current output Active Passive
Terminals
 
Output signal
 
Load RB
 
Source voltage U*
— 
Measuring error
Resolution
Insulation
The current output and digital outputs are electrically 
isolated.
Current output Uco / 32, active
Current output 31 / 32, passive
Permissible source voltage U
q
 for passive outputs in relation to load 
resistance R
B
 where I
max
 = 22 mA. (unreadable type) = Permissible range
A
B
Figure 39  
B
Figure 40  Source voltage for passive outputs
G11596-02
+
-
IE
32-
Uco
32-
IE
31+
R
B
R
B
Uq
31+
Uco
+
-
G10323-02
0
0
100
200
300
400
500
600
5101520253035