64
FEM630 | ELECTROMAGNETIC FLOWMETER  |  OI/FEM630-EN REV. A
Once you have logged on to the corresponding access 
level, you can edit or reset the password. Reset (status 
‘no password defined’) by selecting ‘  ’ as a password.
4  Enter the appropriate password. No password is 
preset in the factory settings. Users can switch to 
the configuration level without entering a 
password. The selected access level remains active 
for 3 minutes. Within this time period you can 
toggle between the process display and the 
configuration level without re-entering the 
password.
5  Use    to confirm the password.
The LCD display now indicates the first menu item on 
the configuration level.
6  Select a menu using   /  .
7  Confirm the selection with   .
8     ...Operation
...Switching to the configuration level 
Read Only
Standard
Reset Password
Back Select
Access Level
Enter Password
**********
 0123456
Next 
Reset password
 0123456
Back
ID
Pin
Password
Select
:     12345678
:     00001
:     *******
Resetting the customer password
If the set password has been forgotten, the password can be 
reset and reassigned.
A one-time password is needed for this purpose and can be 
generated by ABB Service upon request.
To reset the password, the password has to be entered 
incorrectly once for the ‘Standard’ user level. When the 
configuration level is called up again, a new entry ‘Reset 
password’ then appears in the list of access levels.
1  Switch to the configuration level with   .
2  Use   /   to select the ‘Reset password’ entry.
3  Confirm the selection with   .
4  Contact ABB Service and request a one-time password, 
stating the ‘ID’ and ‘Pin’ shown.
5  Enter the one-time password.
Note
The one-time password is only valid once and needs to 
separately requested with each password reset.
6  Confirm the input with   .
After the one-time password has been entered, the password 
for the ‘Standard’ access level is reset and can be reassigned.