70 OI/FEP610/FEH610-EN Rev. B | ProcessMaster FEP610, HygienicMaster FEH610
Error no. /
Range
Error text Current output Digital output Pulse output LCD display Error maskable?
M038.010 /
Electronics
Sensor memory defective.
Mem. or connect. defective.
Replace memory.
Current value - no
change.
No response. Current value - no
change.
Current value - no
change.
Menu "...Group
Masking".
M037.015 /
Electronics
NV chips defect on
Motherboard.
Defective MB. Replace MB.
Call Service.
M032.023 /
Config.
Curr.Out 31 / 32 not
calibrated.
Call Service.
M028.008 /
Config.
Display value is < 1600 h at
Qmax.
Change mass Unit or vol.
Unit for Totalizer.
M026.005 /
Config.
Maintenance interval is
reached.
Perform maintenance.
M024.009 /
Config.
Device not calibrated.
Call Service.
High Alarm or Low
Alarm, depending
on parameter
"Curr.Out at
Alarm".
Current value - no
change.
Current value - no
change.
Current value - no
change.
Menu "...Group
Masking".
Change from one to tw o columns