Table 5: Number and Percent of Results within BG Reference*
BG
Glucose
Level
(mg/dL)
Number
of GM-
Reference
Pairs
Within
±15% /
±15 mg/dL
Within
±20% /
±20 mg/dL
Within
±30% /
±30 mg/dL
Within
±40% /
±40 mg/dL
Outside
±40% /
±40 mg/dL
Overall 11918 66.5 79.4 93.4 97.9 2.1
40-50 152 57.2 66.4 86.8 92.8 7.2
51-80 841 68.4 80.6 93.6 98.1 1.9
81-180 6397 65.7 78.5 92.6 97.4 2.6
181-300 3719 66.5 80.2 94.1 98.7 1.3
301-400 695 71.7 83.5 97.3 99.3 0.7
401-500 114 81.6 90.4 97.4 99.1 0.9
* Comparison to BG was performed using the FreeStyle Precision blood glucose meter. Dierent performance
may be expected when compared to other models of blood glucose meters.