Table 6: Dierence Measures by BG Reference Levels*
BG Reference
Reference Glucose
Level (mg/dL)
Number of GM-
Reference Pairs
Median Absolute
Relative Dierence (%)
Mean Absolute Relative
Dierence (%)
Overall 11918 11.1 13.9
40-50** 152 13.5 19.1
51-80** 841 10.0 12.6
81-180 6397 10.9 13.5
181-300 3719 10.9 12.9
301-400 695 9.1 11.3
401-500 114 7.1 9.2
* Comparison to BG was performed using the FreeStyle Precision blood glucose meter. Dierent performance may be expected when compared to
other models of blood glucose meters.
** For reference values ≤ 80 mg/dL, the mean and median absolute dierences (mg/dL) are presented instead of mean and median absolute relative
dierences (%).
53