Accuracy by Day of Wear
The sensor can be worn for up to 14 days. To show sensor performance over time, the absolute
relative dierence between the System and reference YSI glucose and capillary blood glucose
values (BG) over the 14 day wear is presented in Table 10 and Table 11. The accuracy of GM
versus YSI reference and BG reference was assessed by calculating the percentage of System
readings that were within 15%, 20%, 30% and 40% for reference values 80 mg/dL and above,
and 15 mg/dL, 20 mg/dL, 30 mg/dL and 40 mg/dL for values below 80 mg/dL. The results are
presented in Table 12 for GM vs. YSI reference and in Table 13 for GM vs. BG reference.
Table 10: Dierence Measures by Day (YSI Reference)
YSI Reference
Day
Number of GM-
Reference Pairs
Median Absolute
Relative Dierence (%)
Mean Absolute
Relative Dierence (%)
1 2117 11.4 13.8
2-5 4036 11.0 13.3
6-9 2919 10.2 12.3
10-13 2214 8.4 10.4
14 1037 7.3 9.6
58