Table 13: Number and Percent of Results within BG Reference*
Day Number
of GM-
Reference
Pairs
Within
±15% /
±15 mg/dL
Within
±20% /
±20 mg/dL
Within
±30% /
±30 mg/dL
Within
±40% /
±40 mg/dL
Outside
±40% /
±40 mg/dL
1 1087 61.8 74.1 91.5 96.9 3.1
2-5 4005 64.4 77.6 92.2 97.7 2.3
6-9 3432 64.4 78.2 92.9 97.4 2.6
10-13 2841 71.0 83.4 95.7 98.8 1.2
14 553 80.7 90.2 96.9 99.6 0.4
* Comparison to BG was performed using the FreeStyle Precision blood glucose meter. Dierent performance
may be expected when compared to other models of blood glucose meters.