EasyManuals Logo

Abbott Precision XceedPro User Manual

Abbott Precision XceedPro
79 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
What You See on the Display What You Do Comments
7-4
Questions? Refer to Page 10-9 For Technical Support Instructions.
Chapter 7: Linearity Test
11. Follow the instructions in the CVC kit
package insert. Then apply a drop of the
sample to the target area.
When sufficient sample has been applied, the
monitor beeps, displays Sample Accepted
and automatically starts the test. For a detailed
description of the test strip target area, refer to
the Test Strips section of Chapter 1.
12. Wait for the monitor to analyze the
sample and display the test result.
The monitor counts down then displays
the test result. If an error is detected during
the assay, the assay is terminated and no
result is displayed. Once the assay begins,
all key presses are ignored until the assay is
completed.
13. Note the result. If required, scan or
manually enter the comment code, and
press Enter.
The monitor may be enabled to scan or
enter a 1- to 2-digit comment code. If
there is no prompt to enter a comment
code, skip down to step 14.
The display includes the level and test
numbers, date and time.
Level indicates Level 1-5.
Test indicates replicates 1-4 for that level.
14. You can do one of the following options:
Press 1 – New Level.
Press 2 – Same Level.
Press the Menu button to return to the
Menu Mode.
Press On/Off to turn off monitor.
Remove the test strip from the monitor
and discard it when finished testing.
Follow your facility’s biohazard disposal
policy.
You can press
2 to perform more tests for
the Same Level (return to step 8) or press 1
to perform tests for New Level (return
to step 7).
2.5 mmol/L
2.5 mmol/L

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott Precision XceedPro and is the answer not in the manual?

Abbott Precision XceedPro Specifications

General IconGeneral
BrandAbbott
ModelPrecision XceedPro
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals