Then perform a quality control check with glucose control
solutions and a new test strip. (See “Quality Control
Checks.”) If the control result is within the acceptable
range, review proper testing procedure and repeat your
blood glucose test with a new test strip. If your blood
glucose value is still inconsistent with your previous
results, glucose trends, or how you feel, please contact
your healthcare professional. Follow the advice
of your healthcare professional before you change your
therapy.
Do’s and Don’ts
DO:
• TIGHTLY REPLACE THE CONTAINER LID
IMMEDIATELY AFTER REMOVING A TEST STRIP.
•
Make sure the code number on the meter matches
the code number on the test strip container label.
• Keep the meter clean.
• Store the test strips in the original container.
DON’T:
• Use expired test strips.
•
Use a code chip from another blood glucose system.
•
Store the meter or test strips in extreme temperatures.
• Store the meter in damp or humid areas.
• Insert any object other than a test strip into the test
strip slot.
• Remove or insert a code chip while meter is on.
• Test with the meter placed on hot or cold surfaces.