37
If blood has entered the meter, wipe the area around
the test strip slot and the display screen. Allow areas
you have wiped time to dry thoroughly.
The meter, including the test strip slot and display
screen, must be thoroughly cleaned and disinfected
at least once a day after use. Carefully check all
recesses, grooves and gaps.
Suitable Disinfecting/Cleaning Solutions
Products suitable for use are:
70% Isopropyl Alcohol
Mild washing up liquid mixed with water
10 % household bleach solution (1 part bleach plus
9 parts water) made same day