Construction Hand held
Protection Class III
Rated Operation Continuous operation
Data Port 3-wire RS-232C Serial Port
Electromagnetic Compatibility
This meter meets the electromagnetic immunity
requirements as per ISO 15197 Annex A. The chosen
basis for electrostatic discharge immunity testing
was basic standard IEC 61000-4-2. In addition, it
meets the electromagnetic emissions requirements
as per EN 61326. Its electromagnetic emission is
thus low. Interference from other electrically driven
equipment is not to be anticipated.
Performance Analysis
The performance data for the Accu-Chek Advantage
System (Accu-Chek Advantage Meter with
Accu-Chek Advantage Plus Test Strips) were
obtained using capillary blood from diabetic patients
(method comparison, accuracy), venous blood
(repeatability) and control solution (reproducibility).
The system is calibrated with venous blood
containing various levels of glucose. The reference
values are obtained using the hexokinase method.
For method comparison, the results were compared
with results obtained using the hexokinase method
with deproteinization (automatic analyzer). The
hexokinase method is traceable to an NIST standard.
56