101
The year, time and date are not
set. – Set time and date. See
Chapter 1, “Setting the Year,
Time and Date.”
The test drum in the meter is
empty. A dot also appears in the
test strip count window on the
back of the meter. – Insert a new
test drum. If the message is still
displayed, call the ACCU-CHEK
Customer Care Service Center at
1-800-858-8072.
The bar code information
stored in the meter is lost.
Slide and release
to turn
off the meter. Open and close
the drum door. The next time
you slide and release
to
turn the meter on and perform
a test, the drum symbol is
displayed and the bottle
symbol ashes. The 90-day
use-by period is no longer
calculated for this test drum.
The bar code on the test drum
is damaged. You can no longer
perform tests with test strips
from this drum. – Insert a new
drum.