Cleaning and Disinfecting the Lancing Device
To clean and disinfect without damaging the lancing device, follow these procedures carefully:
When to Clean and Disinfect the Lancing Device
Clean the lancing device to remove visible dirt or other material prior to disinfecting.
Clean and disinfect the lancing device at least once per week to remove visible dirt or other material
for safe handling.
Clean and disinfect the lancing device before allowing anyone else to handle the lancing device, for
instance, if you have someone assisting you. Do not allow anyone else to use the lancing device.
NOTE
Do not throw away the lancing device cap after each use. Use approved cleaning and disinfecting
products on it.
You should always remove the lancet before cleaning or disinfecting the lancing device.
Approved Cleaning and Disinfecting Product
The following product has been approved for cleaning and disinfecting the lancing device:
Super Sani-Cloth (EPA* reg. no. 9480-4)
* Environmental Protection Agency
Super Sani-Cloth can be purchased from the following sources:
Amazon.com
Ofcedepot.com
Walmart.com
man_07699336001_01_EN.indb 88 28.09.2015 13:26:58

Table of Contents

Other manuals for Accu-Chek Compact Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact Plus and is the answer not in the manual?

Accu-Chek Compact Plus Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1.5 µL
Test Time5 seconds
Automatic CodingYes
DisplayLCD
Measurement MethodElectrochemical
Battery Type2 x AAA
Battery LifeApprox. 1000 tests
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Operating Conditions10-40°C, 10-90% RH