1. Wash hands thoroughly with
soap and water.
2. Wipe the entire lancing device
surface with the approved
cleaning and disinfecting
product. Always use the same
product for both cleaning and
disinfecting.
Carefully wipe each groove
with the approved cleaning and
disinfecting product.
3. Remove and wipe the inside of
the lancing device cap.
WARNING
The effect of using more than one product interchangeably to clean and disinfect the lancing device
has not been tested. Always use Super Sani‑Cloth (EPA reg. no. 9480‑4) to clean and disinfect the
lancing device. Do not use any other cleaning or disinfecting solutions. Using solutions other than the
Super Sani‑Cloth could result in damage to the lancing device.
man_07699336001_01_EN.indb 90 28.09.2015 13:27:00

Table of Contents

Other manuals for Accu-Chek Compact Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact Plus and is the answer not in the manual?

Accu-Chek Compact Plus Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1.5 µL
Test Time5 seconds
Automatic CodingYes
DisplayLCD
Measurement MethodElectrochemical
Battery Type2 x AAA
Battery LifeApprox. 1000 tests
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Operating Conditions10-40°C, 10-90% RH

Related product manuals