125
Acoustic Mode
6
mg/dL meters
First the hundreds are announced, then the tens, and finally the units.
There is a short pause between each group of beeps. The hundreds are
always announced, even when the result is below 100. The hundreds in
this case are represented by 1 long beep, signifying zero.
Examples:
The result 180 mg/dL is announced as follows:
1 beep – pause – 8 beeps – pause – 1 long beep
( – – – – – – – – – — )
The result 72 mg/dL is announced as follows:
1 long beep (for 0 hundreds) – pause – 7 beeps – pause – 2 beeps
( — – – – – – – – – – )
Results that are lower than 0.6 mmol/L or 10 mg/dL and are displayed
as
are announced acoustically as 0 – decimal point – 0 and
0 – 0 – 0, respectively.
Results that are higher than 33.3 mmol/L or 600 mg/dL and are
displayed a
are announced acoustically as 9 – 9 – decimal point – 9
and 9 – 9 – 9, respectively.
man_06919570001_01_EnCan.indd 125 18.01.2013 18:38:45