Checking your meter
Farbe: P 249 – Prüfmittelnummer: 203
Subsequently hold the meter directly above a waste bin with the test
strip pointing downwards.
Press the Test button to turn the meter off.
The used test strip drops out. The test result is saved together with the
bottle symbol
.
Before the meter turns off, it shows you the number of test strips
remaining in the drum (see Chapter 2.3.2).
Wipe the tip of the dropper with a clean, dry paper towel and close
the bottle tightly.
Notes
i
The control solution you apply to a washable object must come
from a suspended drop. Do not use the drop if it has run down
the outside of the tip of the dropper. This may lead to results that
are too high.
i
Control solution that is to be drawn up by the test strip must just
have been freshly applied to the object. The control solution must
not be left exposed on the surface for any length of time, or it
may give results that are too high.
i
Results flagged as control tests with the bottle symbol are not
included in the calculation of averages and lowest and highest
results.
man_06919570001_01_EnCan.indd 136 18.01.2013 18:38:50

Table of Contents

Other manuals for Accu-Chek Compact Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact Plus and is the answer not in the manual?

Accu-Chek Compact Plus Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1.5 µL
Test Time5 seconds
Automatic CodingYes
DisplayLCD
Measurement MethodElectrochemical
Battery Type2 x AAA
Battery LifeApprox. 1000 tests
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Operating Conditions10-40°C, 10-90% RH