183
Symbols, error messages and troubleshooting
12
If this happens … it means:
The meter is turned on,
but the display is blank.
The display is defective.
Please call Accu-Chek Customer Care at
1-800-363-7949.
The clock has stopped or
the clock is slow.
The meter was exposed to a temperature
below –10 °C and the batteries are
beginning to freeze.
Turn the meter off. Move to a place where
the temperature is between +10 and
+40 °C and wait for the temperature of
the meter to adjust to this temperature.
When the meter is turned
on, the date is shown as
0-00 and the time is
shown as 0:00.
The meter was exposed to a
temperature below –10 °C and the
batteries are beginning to freeze.
Press the Test button to turn the meter
off. Move to a place where the
temperature is between +10 and
+40 °C and wait for the temperature
of the meter to adjust to this
temperature.

man_06919570001_01_EnCan.indd 183 18.01.2013 18:39:03