Symbols, error messages and troubleshooting
Farbe: P 249 – Prüfmittelnummer: 203
Error message Meaning

The year, time and date have not been
set.
Set these as described in Chapter 3.
The test strip drum in the meter is empty.
Insert a new test strip drum.
The bar code information from the test
strip drum has been lost.
Press the Test button to turn the meter
off. Open and close the drum
compartment cover. The meter rotates
the drum and reads the bar code.
When you next turn on the meter to
perform a test, the drum symbol and
flashing bottle symbol are displayed.

man_06919570001_01_EnCan.indd 184 18.01.2013 18:39:04