No. Meaning of the alert What you have to do Page
A1 Cartridge Low Warning Silence and confirm the alert. 136
Change the cartridge before it is
completely empty.
A2 Low PowerPack Warning Silence and confirm the alert. 137
Replace the PowerPack within the next
24 hours.
A3 Set Time/Date Silence and confirm the alert. 137
Put your insulin pump into RUN.
Set the time and date in the SETUP menu.
A4 Cartridge/Adapter Alert Silence and confirm the alert. 138
Act according to section 7.1.4
Alert A4: Cartridge/Adapter Alert”.
A5 End of Use Time Alert Silence and confirm the alert. 143
(t1 or Contact your Accu-Chek customer service
t2) center for insulin pumps to ensure
continuation of your insuli pump therapy.
A6 Temporary Basal Silence and confirm the alert. 143
Rate Cancelled Put your insulin pump into RUN if necessary.
Ensure that the cancellation was intended
and program a new temporary basal rate if
necessary.
A7 Temporary Basal Silence and confirm the alert. 144
Rate Over Decide whether a further temporary basal
rate change is appropriate and program
one if necessary.
A8 Bolus Cancelled Silence and confirm the alert. 144
Put your insulin pump into RUN if necessary.
Ensure that the cancellation was intended
and program a new bolus if necessary.
No. Meaning of the error What you have to do Page
E1 Empty Cartridge Silence and confirm the error. 146
Act according to section 7.2.1
“Error E1: Empty Cartridge”.
When finished, put your insulin pump
into RUN.
E2 PowerPack Depleted Silence and confirm the error. 147
Act according to section 7.2.2
“Error E2: PowerPack Depleted”.
When finished, put your insulin pump
into RUN.
E3 Automatic Off Silence and confirm the error. 147
Put your insulin pump into RUN.
E4 Occlusion Silence and confirm the error. 148
Act according to section 7.2.4
“Error E4: Occlusion”.
When finished, put your insulin pump
into RUN.
E5 End of Use Time Silence and confirm the error. 150
(t1 or Act according to section 7.2.5
t2) “Error E5: End of Use Time”
E6 Mechanical Error Silence and confirm the error. 150
Act according to section 7.2.6
“Error E6: Mechanical Error”.
E7 Electronic Error Act according to section 7.2.7 151
“Error E7: Electronic Error”.
Overview of the alerts Overview of the errors