Accu-Chek D-TRONplus insulin pump Reference Manual
33
Accu-Chek D-TRONplus insulin pump Reference Manual
32
2.4 Operating logic and overview of
Accu-Chek D-TRONplus insulin pump functions
The operation of your Accu-Chek D-TRONplus insulin pump is logical and simple.
Most operations are carried out using the following four steps:
1. Select a function by pressing
d
until the corresponding symbol appears on the
display.
2. Confirm the selection by pressing
f
.
Info: If
f
is not pressed within 20 seconds*, your insulin pump auto-
matically returns to STOP or RUN (timeout).
3. Set/scroll the selected value (flashing) by pressing
a
and/or
s
.
Info: If no key is pressed within 20 seconds*, your insulin pump auto-
matically returns to STOP or RUN (timeout).
The changes you just made are not saved.
4. Confirm the changes by pressing
f
.
Info: If
f
is not pressed within 20 seconds*, your insulin pump auto-
matically returns to STOP or RUN (timeout).
The changes you just made are not saved.
* This parameter may vary based on the standard configuration of your insulin pump
and changes made using the Accu-Chek DiaLog Pump Programming tool.
2.3 Operating status (STOP/RUN)
The operating status of your Accu-Chek D-TRONplus insulin pump may be either
STOP or RUN.
RUN
During normal use, your insulin pump is in RUN and insulin is
continuously delivered (basal rate). A bolus can be programmed
when needed. The time, the symbol for the current basal rate
profile, and the value of the current basal rate or, if activated, in-
formation on special functions (e.g. extended bolus, temporary
basal rate or KeyLock) appear on the display when your insulin
pump is in RUN.
STOP
In STOP, your insulin pump does not deliver insulin. For exam-
ple when you return the piston rod, change the cartridge, con-
nect the adapter and/or infusion set or prime your infusion set,
your insulin pump must be in STOP. The time, the STOP symbol
and -S-- appear on the display when your insulin pump is in
STOP (if activated, the KeyLock symbol is also displayed).
In both STOP and RUN, a blinking colon in the time display indicates that your in-
sulin pump is functioning.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek D-TRONplus and is the answer not in the manual?

Accu-Chek D-TRONplus Specifications

General IconGeneral
BrandAccu-Chek
ModelD-TRONplus
CategoryMedical Equipment
LanguageEnglish

Related product manuals