Accu-Chek D-TRONplus insulin pump Reference Manual
121
Accu-Chek D-TRONplus insulin pump Reference Manual
120
To adjust the beep volume
1. Make sure your insulin pump is in RUN and the KeyLock is
deactivated or the keys are unlocked.
2. Press
d
9 times to select the SETUP menu.
3. Confirm with
f
.
4. Press
d
twice to move to the beep volume function.
5. Confirm with
f
.
The currently programmed beep volume flashes.
6. Press/scroll
a
or
s
to increase or decrease the beep volume.
7. Confirm with
f
.
Your insulin pump returns to RUN.
Info: If
f
is not pressed within 20 seconds*, your insulin pump automa-
tically returns to RUN (timeout).
The changes you just made are not saved.
* This parameter may vary based on the standard configuration of your insulin pump
and changes made using the Accu-Chek DiaLog Pump Programming tool.
6. Press/scroll
a
to increase or
s
to decrease the duration for
the automatic off in 1 hour intervals.
7. If you would like to set the automatic off function to OFF,
press/scroll
s
until OFF appears in the display.
8. Confirm with
f
.
Your Accu-Chek D-TRONplus insulin pump returns to RUN.
Info: If
f
is not pressed within 20 seconds*, your insulin pump automa-
tically returns to RUN (timeout).
The changes you just made are not saved.
6.5 Adjust the beep volume
Your insulin pump beeps when a key is pressed or when an alert or error occurs.
The beep volume is set to level 2 (normal)* in the standard configuration. This beep
volume can be adjusted in the SETUP menu or with the Accu-Chek DiaLog Pump
Programming tool.
There are five volume levels:
0 no beeps (beeps are silenced) 3 loud
1 low 4 very loud
2 normal
Info: All beeps can be turned off with the exception of safety-relevant
beeps such as the ones for alerts and errors. For safety reasons, we recom-
mend that you never turn off the volume completely.
* This parameter may vary based on the standard configuration of your insulin pump
and changes made using the Accu-Chek DiaLog Pump Programming tool.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek D-TRONplus and is the answer not in the manual?

Accu-Chek D-TRONplus Specifications

General IconGeneral
BrandAccu-Chek
ModelD-TRONplus
CategoryMedical Equipment
LanguageEnglish

Related product manuals