6.4 Automatic off
This safety function triggers an “Error E3: Automatic Off” (your insulin pump stops
insulin delivery!) if no key is pressed within a certain time.
Automatic off is set to OFF* in the standard configuration. It can either be set to
OFF or programmed for up to 24 hours in 1 hour intervals in the SETUP menu or with
the Accu-Chek DiaLog Pump Programming tool.
To program the automatic off
1. Make sure your Accu-Chek D-TRONplus insulin pump is in
RUN and the KeyLock is deactivated or the keys are unlocked.
2. Press
d
9 times to select the SETUP menu.
3. Confirm with
f
.
4. Press
d
once to move to the automatic off function.
5. Confirm with
f
.
The parameter for the automatic off flashes.
* This parameter may vary based on the standard configuration of your insulin pump
and changes made using the Accu-Chek DiaLog Pump Programming tool.
Info: If
f
is not pressed within 20 seconds*, your insulin pump automa-
tically returns to RUN (timeout).
The previously entered temporary basal rate remains unchanged and active.
Caution: Your insulin pump confirms the cancellation and an “Alert A6:
Temporary Basal Rate Cancelled” occurs. Silence and confirm the alert by
pressing
f
twice.
Please refer to section 7.1.6 “Alert A6:Temporary Basal Rate Cancelled” for
further information.
Option B
1. Put your insulin pump into STOP by pressing
d
twice
and
f
once.
Caution: Your insulin pump confirms the cancellation and an “Alert A6:
Temporary Basal Rate Cancelled” occurs. Silence and confirm the alert by
pressing
f
twice. Put your insulin pump into RUN (press
d
and
f
) to con-
tinue the insulin delivery.
Please refer to section 7.1.6 “Alert A6:Temporary Basal Rate Cancelled” for
further information.
If an extended bolus is active at the same time, it is also cancelled when
your insulin pump is put into STOP. An “Alert A8: Bolus Cancelled” occurs
first to warn you that the extended bolus has also been cancelled.
Press
f
once to silence the “Alert A8: Bolus Cancelled”. Only after silen-
cing the“Alert A6:Temporary Basal Rate Cancelled”appears on the display.
Silence and confirm it by pressing
f
twice. The history of alerts and errors
records both alerts.
Please refer to sections 6.1.3 “Review the history of alerts and errors”and
7.1.8 “Alert A8: Bolus Cancelled” for further information.
* This parameter may vary based on the standard configuration of your insulin pump
and changes made using the Accu-Chek DiaLog Pump Programming tool.