EasyManuals Logo

Accu-Chek Guide User Manual

Accu-Chek Guide
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
96
Meter and Lancing Device Cleaning and Disinfecting
8
NOTE
For technical assistance or questions on cleaning and disinfecting, contact the Accu‑Chek Customer
Care Service Center at 1‑800‑858‑8072.
Cleaning and Disinfecting the Meter
WARNING
If the meter is being operated by a second person who is providing testing assistance to the user, the
meter and lancing device should be cleaned and disinfected prior to use by the second person.
To clean and disinfect without damaging the meter, follow these procedures carefully.
When to Clean and Disinfect the Meter
Clean the meter to remove visible dirt or other material prior to disinfecting.
Clean and disinfect the meter at least once per week and when blood is present on the surface of the
meter.
Clean and disinfect the meter before allowing anyone else to handle the meter. Do not allow anyone else
to use the meter on themselves for testing purposes.
NOTE
Using cleaning and disinfecting products could result in damage to the meter. If you notice any of the
following signs of deterioration after cleaning and disinfecting your meter, stop using your meter and
contact the Accu‑Chek Customer Care Service Center at 1‑800‑858‑8072: residue around buttons,
clouding of display, button malfunction, out‑of‑range control results.
What to Clean and Disinfect
The following parts of the meter should be cleaned and disinfected:
The area around slots and openings (do not get any moisture in slots or openings)
The meter display
The entire meter surface
54988_08100918001_EN.indb 96 9/30/16 5:04 PM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Guide and is the answer not in the manual?

Accu-Chek Guide Specifications

General IconGeneral
Measurement Range10 to 600 mg/dL
Test TimeLess than 4 seconds
Battery LifeApproximately 1000 tests
ConnectivityBluetooth
Sample Size0.6 μL
WeightApprox. 40 g (1.4 oz) with battery
Battery Type2 x CR2032 lithium coin cells
DisplayLCD
Operating Conditions10-40°C (50-104°F), 10-90% relative humidity
Storage Conditions-25°C to +70°C (-13°F to +158°F)