OMNIVERSA
®
MULTI-MODALITY THERAPY USER MANUAL - 65 -
COPYRIGHT © 2017-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Pain Control
HVPC Sensory
Stimulates segmental pain modulation and vasodilation, increasing local circulation. Application of negative polarity
over the edematous site has been demonstrated to reduce post-traumatic edema in clinical trials.
Intensity: Set amplitude to as high as comfortably tolerated. Increase amplitude during the treatment in response to
accommodation to the electrical stimulation, when and as needed.
Adjustable
Parameter
(Y/N)
Parameter
Name
Range (unit)
Default (unit)
Y PULSE RATE 1…200 (Hz) 100 (Hz)
Y
TREATMENT
TIME
0…60 (min) 15 (min)
IFC - PREMOD Sensory
This program is used for symptomatic relief of superficial and deep pain from localized dermatome or segmental
origin. Duration of relief is typically several hours. Fast onset of relief usually happens within 15-20 minutes. Target
tissue is superficial and deep. Bilateral, Bipolar, or Quadripolar application; through the painful area; or over the
involved spinal segments. Apply parallel to incision line for post-operative pain management. The intensity should be
set to elicit a pleasant tingling sensation, just below a muscle contraction.
User
Adjustable
Parameter
(
Y/N
)
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y CARRIER FREQ 2…10 KHz 5.0 (kHz)
N BURST FREQUENCY 100 (Hz)
N BURST RATE SCAN 20 (%)
N RATE SCAN ON
N RATE SCAN TIME 20 (s)
Y VECTOR
Slow 45, Fast 45, Slow 90,
Fast 90, OFF
OFF
Y TREATMENT TIME 0…60 (min) 15 (min)