- 92 - OMNIVERSA
®
MULTI-MODALITY THERAPY USER MANUAL
COPYRIGHT © 2017-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
OMNIVERSA PROGRAM MENU - COMBINATION THERAPY
COMBINATION THERAPY INDICATIONS
Pain
1 POST TRAUMATIC - POST SURGICAL
2 ACUTE
3 CHRONIC
Increase Local Circulation
1 HVPC SENSORY
Decrease Muscle Tone - Spasm
1 PREMOD MOTOR
Combination Therapy Manual Mode
1 HVPC Waveform – Thermal Ultrasound
2 HVPC Waveform – Pulsed/Continuous Ultrasound
3 PREMOD Waveform – Thermal Ultrasound
4 PREMOD Waveform – Pulsed/Continuous Ultrasound
PAIN MANAGEMENT – INCREASE LOCAL CIRCULATION
COMBO: Post Traumatic – Post Surgical & Increase Local Circulation
HVPC Sensory
This protocol combines subthermal therapeutic ultrasound with high volt pulsed current (HVPC) electrical stimulation
current at a sensory (100 Hz) frequency for 1) superficial pain reduction and 2) increase local circulation to targeted
tissue area. Sensory current stimulates vasodilation, increasing local circulation. Application of negative polarity over
the edematous site has been demonstrated to reduce post-traumatic edema in clinical trials. In addition, if pain is
present, can reduce it via segmental pain modulation mechanisms. Therapeutic ultrasound can resolve post-traumatic
edema, reduce pain and facilitate tissue healing.
Duration of relief is typically for several hours. Fast onset of relief usually happens within 15-20 minutes. Target
tissue is superficial and deep. Application; Monopolar, using channel B, place the red (positive) electrode adjacent to
the treatment area. The ultrasound transducer is the black (-) electrode. The intensity should be set to elicit a pleasant
tingling sensation, just below a muscle contraction.
Adjustable
Parameter
(Y/N)
Parameter Name
Range (unit)
Default (unit)
Y PULSE RATE 1…200 (Hz) 125 (Hz)
Y TREATMENT TIME 0…60 (min) 15 (min)
Adjustable
Parameter
(Y/N)
Parameter Name
Range (unit)
Default
units
Y ULTRASOUND
FREQUENCY
1 or 3 MHz 3 (MHz)
Y PULSED -
CONTINUOUS
CW, 5,10, 20,25, 33, 50 (%) 20 (%)
Y TRANSDUCER SIZE 0.8 cm
2
or 5 cm
2
5 cm
2
Y TREATMENT TIME 0…60 (min) 15 (min)