IFC – PREMOD Sequential
This protocol provides symptomatic relief of superficial and deep pain and pain of local, generalized single or
multi-segmental nature. Combines sensory and motor stimulation in 3 sequences of 4 Hz, 15 Hz and 100 Hz
repeated every 15 seconds. Duration of relief is typically from 2 to 6 hours. Rapid onset of relief usually within
5-15 minutes. Target tissue is superficial and deep. Bipolar placement over local and distal trigger points.
Quadripolar placement over area of local pain, or at involved spinal segment. The intensity should be set to elicit
a moderate muscle twitch. The protocol consists of three sequences. This sequence is repeated throughout the
program until the time runs out. The amplitude does not change when the sequence shifts in the protocol
.
Adjustable Parameters
Adjustable
Parameter
(Y/N)
Parameter
Name
Adjustable Range (unit) Default (unit)
Y
VECTOR
Off, Slow 45, Fast 45, Slow 90, Fast 90, Off
Y
TREATMENT TIME
0…60 (min) 15 (min)
Note:
Total treatment time on the timer is set at 15 minutes. At the end of treatment timer, the output shuts off regardless
of sequence completion
.
IFC – PREMOD Sequential, First Sequence
Adjustable
Parameter
(Y/N)
Parameter Name Adjustable Range (unit)
Default
(unit)
N CARRIER FREQ 5.0 (kHz)
N
BURST FREQ
4
N RATE SCAN TIME 15
N RATE SCAN % 50 (%)
Y VECTOR Off, Slow 45, Fast 45, Slow 90, Fast 90 Off
N TREATMENT TIME (SEQ) 15 (sec)
IFC – PREMOD Sequential, Second Sequence
Adjustable
Parameter
(Y/N)
Parameter Name Adjustable Range (unit)
Default
(unit)
N CARRIER FREQ 5.0 (kHz)
N BURST FREQ 15
N RATE SCAN TIME 15
N RATE SCAN % 50 (%)
Y VECTOR Off, Slow 45, Fast 45, Slow 90, Fast 90 Off
N TREATMENT TIME 15 (sec)