EasyManua.ls Logo

ACP OMNIVERSA - Program Details: Neuro Muscular Re-Education; UE BIPHASIC PENS Protocol

ACP OMNIVERSA
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 72 - OMNIVERSA
®
MULTI-MODALITY THERAPY USER MANUAL
COPYRIGHT © 2017-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Neuro Muscular Re-Education
UE BIPHASIC PENS
This program is used to re-educate upper extremity agonist - antagonist muscle pairs to enhance reciprocal and
fast movement. It is called biphasic because the agonist fires first followed by the antagonist. Due to reciprocating
action of the muscle groups there is no third phase. Functional task simulation is reciprocal movement such as
typing, boxing, piano playing, swimming, upper extremity positioning speed walking, rowing, etc. Patterned
Electrical Neuromuscular Stimulation (PENS) is a precise reproduction of EMG firing patterns of muscle agonist and
antagonist pairs. PENS should be used prior to Therex to improve recruitment and motor performance. It’s also used to
decrease spasticity and muscle tone.
User
Adjustable
Parameter
(
Y/N
)
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y PHASE DURATION 40…100 (μs) 70 (μs)
N PULSE RATE 50 (Hz)
Y TREATMENT TIME 0…60 (min)
15 (min)
Y CYCLE TIME 2.5; 1.5; .75; .6; .5 (sec) .75 (sec)
N PATTERN Bi Phasic
For unilateral involvement, apply CH A to the most involved side
For Torticollis, place CH A on the side that the head is tilted toward