EasyManua.ls Logo

ACP OMNIVERSA - Program Details: Muscle Disuse Atrophy; MFAC LARGE MUSCLE Protocol

ACP OMNIVERSA
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OMNIVERSA
®
MULTI-MODALITY THERAPY USER MANUAL - 75 -
COPYRIGHT © 2017-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Treatment of Muscle Disuse Atrophy
MFAC LARGE MUSCLE
Program used for muscle re-education and treatment of muscle disuse atrophy. Activation of muscle with electrical
stimulation at a high intensity for a short time with long “ON/OFF ramps reduces atrophy. The patient should
participate to the extent possible by contracting during the stimulation “ON” time. Target tissues are superficial and
deep muscle groups. Bipolar set-up over agonist and antagonist muscles for target muscle. The intensity should be set
to elicit a grade 2 to 4 muscle contraction. Treatment time should be reduced based on muscle fatigue. The affected joint
movement should be restricted by holding manually, or using weights or exercise bands.
User
Adjustable
Parameter
(
Y/N
)
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y CARRIER FREQ 2.0…10.0 (kHz) 2.5 (kHz)
Y BURST FREQUENCY 0.0…200 (Hz) 75 (Hz)
N BURST RATE SCAN 20 (%)
N RATE SCAN ON
N RATE SCAN TIME 4 (s)
Y TREATMENT TIME 0…60 (min) 10 (min)
Y 2 B OUTPUT DELAY 0.0…9.9 (s) 0.0 (s)
Y RAMP UP 0.0…9.9 (s) 2.0 (s)
Y RAMP DOWN 0.0…9.9 (s) 2.0 (s)
Y T-ON 0…60 (s) 10 (s)
1
Y T-OFF 0…120 (s) 50 (s)
1
Y CHANNEL MODE
A/B SIMULTANEOUS, A/B ALTERNATE, B
DELAYED
A/B SIM.
1
When T-OFF = 0, T-ON = CONTINUOUS.
2
B OUTPUT DELAY parameter is shown only when the Channel Mode is set to B DELAYED.