- 80 - OMNIVERSA
®
MULTI-MODALITY THERAPY USER MANUAL
COPYRIGHT © 2017-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
MFAC – LVPC HAND CONTROL MANUAL
Program provides for active assisted treatment to develop functional movement of superficial and deep muscle.
Activation of muscle with electrical stimulation occurs during functional movement when the therapist presses and
holds the start button. The patient should participate to the extent possible with the functional movement by contracting
during the stimulation “ON” time. Target tissue is superficial and deep muscle groups. Bipolar set-up over agonist and
antagonist muscles for target muscle. Set intensity to elicit a grade 3-5 muscle contraction. Reduce treatment time based
on muscle fatigue.
User
Adjustable
Parameter
Y/N
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y CARRIER FREQ 2.0, 2.5, 4.0, 5.0, 10.0 (kHz) 2.5 (kHz)
Y BURST FREQUENCY 0.0…250 (Hz) 50 (Hz)
N BURST RATE SCAN 20 (%)
N RATE SCAN ON
N RATE SCAN TIME 4 (s)
Y TREATMENT TIME 0…60 (min) 15 (min)
N CHANNEL MODE A/B SIM.
MFAC- LVPC HAND CONTROL SEQUENCED
Program provides for active assisted treatment to develop functional movement of superficial muscle. Activation of
muscle with electrical stimulation during functional movement. The patient should participate to the extent possible
with the functional movement by contracting during the stimulation “ON” time. Target tissue is smaller superficial
muscle groups. Bipolar set-up over agonist and antagonist muscles for target muscle. Set intensity to elicit a grade 3-5
muscle contraction. Reduce treatment time based on muscle fatigue.
User
Adjustable
Parameter
Y/N
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y PHASE DURATION 40…400 (μs) 70 (μs)
Y PULSE RATE 0.0…250 (Hz) 50.0 (Hz)
Y Ron, Roff, Ton 1,1,5
N MODULATION OFF
N RATE SCAN TIME 4 (s)
Y TREATMENT TIME 0…99 (min) 15 (min)
N CHANNEL MODE AB SImul …Alterating … Delayed A/B SIM.