MFAC Tone - Spasm
This program provides for stimulation of the spastic muscle. The short rest time exhausts the muscle metabolism
reducing spasm- tone. Slow ramps decrease the potential to trigger spasm. Intensity should be set to elicit a grade
2 to 3 muscle contraction Treatment time can be reduced based on muscle fatigue and inhibition of spasticity.
User
Adjustable
Parameter
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y CARRIER FREQ 2, 2.5,3,3.5,4, 5, 6, 7, 8, 9, 10 kHz 5.0 (kHz)
Y BURST FREQUENCY 1…100 (Hz) 90 (Hz)
Y BURST RATE SCAN 0…50 (%) 20 (%)
Y RATE SCAN ON, OFF ON
Y RATE SCAN TIME 0…20 (s) 4 (s)
Y TREATMENT TIME 0…60 (min) 15 (min)
Y B OUTPUT DELAY 0.0…80 (s) 0.0 (s)
Y RAMP UP 0.0…9 (s) 2.0 (s)
Y RAMP DOWN 0.0…9 (s) 2.0 (s)
Y T-ON 0…60 (s) 10 (s)
Y T-OFF 0…120 (s) 10 (s)
Increase Range of Motion (ROM)
MFAC INCREASE ROM
Program used for stimulation of range of motion through muscle activation of the agonist or agonist-antagonist
muscle pairs. Activation of muscle with electrical stimulation at a moderate intensity for a longer time (Endurance
protocol) with longer “ON/OFF” ramps improves range of motion. The patient should participate to the extent
possible by contracting during the stimulation “ON” time. Bipolar set-up over agonist and antagonist muscles for
target muscle. The intensity should be set to elicit a grade 2 to 3 muscle contraction. Treatment time should be
reduced based on muscle fatigue.
Adjustable
Parameter
(Y/N)
Parameter Name
Range (unit)
Default (unit)
Y CARRIER FREQ 2, 2.5,3,3.5,4, 5, 6, 7, 8, 9, 10 kHz 2.5 (kHz)
Y BURST FREQUENCY 1…100 (Hz) 35 (Hz)
Y BURST RATE SCAN 0…50 (%) 20 (%)
Y RATE SCAN ON, OFF ON
Y RATE SCAN TIME 0…20 (s) 4 (s)
Y TREATMENT TIME 0…60 (min) 15 (min)
Y B OUTPUT DELAY 0.0…80 (s) 0.0 (s)
Y RAMP UP 0.0…9 (s) 2.0 (s)
Y RAMP DOWN 0.0…9 (s) 2.0 (s)
Y T-ON 0…60 (s) 4 (s)
Y T-OFF 0…120 (s) 12 (s)