Possible causes Solutions
The Mini LED does not produce enough power
Check the power. Please read the chapter Check the power
page 23
The end of the optical guide is too far from the cure site
Place the end of the optical guide at 2 mm from the cure
site
10.5 Malfunction of the power tester
Possible causes Solutions
Defective light indicator Send the Mini LED to Acteon's Customer Services team.
The optical guide is defective, damaged or dirty
Clean the optical guide using the multi-purpose syringe
air function and/or return the Mini LED to Acteon's
Customer Service team
The reflector is defective or dirty
Clean the reflector using the multi-purpose syringe air
function and/or return the Mini LED to Acteon's Customer
Service team
The power tester window is defective or dirty
Clean the power tester window and/or return the Mini LED
to Acteon's Customer Service team
10.6 Other malfunctions
If the Mini LED is not working for any other reason, contact the Acteon Customer Services Team.
If you need to return your Mini LED, please ensure the optical guide and battery are suitably packed to prevent any
impact damage during transportation.
Keep the original packaging of the medical device and use it to return the device for servicing, maintenance or repair.