Capacity 2500 mA / hr
Output voltage 3.7 V - 4.2 V
Diameter maximum 23 mm
length 99 mm
Weight 74 g
11.6 Base
Supply voltage 12 V DC
Protection 3 A / 125 V AC fuse
Ingress protection rating IPX0
11.7 Environmental characteristics
Operating temperature +10°C to +30°C
Operating RH 30% to 75%
Atmospheric operating pressure 800 hPa to 1060 hPa
Maximum operating altitude Equal to or less than 2000 metres
Storage temperature 0°C to +50°C
Storage RH 10% to 95%, including condensation
Atmospheric storage pressure Between 500 hPa and 1060 hPa
11.8 Environmental restrictions
Usage premises
Usable in all medical premises. The medical device must not be used
in an operating theatre or outdoors.
Use in gas-filled atmosphere
The medical device is not designed for use in a type AP or APG gas-
filled atmosphere or in the presence of anaesthetic gases.
Immersion The handpiece must not be immersed.
11.9 Main performance characteristics
Wavelength between 440 and 460 nm.
Irradiance of 1125mW/cm
2
to 2400mW/cm2, calculated based on the active diameter of 6.8mm with the opalescent
optical guide.
Irradiance of 785 mw / cm
2
- 1680 mW / cm
2
, calculated based on the active diameter of 6.8 mm with the amber-
coloured optical guide.
Page 28/38 - User Manual | Mini LED™ | J02541|V8|(07)| 01/2017|NG13EN010H