OPERATOR’S MANUAL • X-Mind AC • (06) • 05/2018• NXACEN010F
!
WARNING
This modication brought to the exposure time is momentary and it will be lost unless it is saved. (refer to Chapter 6) To restore the pre-
vious values, press one of the keys with the LED turned off on the control panel.
7° - MAKE THE EXPOSURE
1. Take the control button of the timer relevant to the selected tubehead and keep a safety distance (at least 2
meters) from the tubehead, in order to be able to constantly check the radiographic exposure
2. Advise the patient to remain still
3. On the control button press the X-ray key and keep it pressed until the acoustic signal (beep) stops
and LED yellow turns off
PLEASE NOTE
If the “X-ray” key is released earlier than the selected exposition time, the exposure is immediately interrupted and the E12 error message
appears on the display.
4. At the end of the exposure, the green LED PAUSE intermittently turns on
5. The display indicates the actual exposure time
6. All the timer functions are inhibited
PLEASE NOTE
The pause time is necessary to allow the X-ray tube to cool down.
This time is calculated by the microprocessor, depending on the exposure time, with a ratio of 1:32 (32 s of pause are required for each
second of exposure)
A NEW EXPOSURE WILL BE POSSIBLE AFTER THE GREEN LED HAS TURNED OFF
REPEAT THE OPERATIVE SEQUENCE FROM POINT 2 TO POINT 7 TO MAKE A NEW EXPOSURE