EasyManua.ls Logo

AED HeartPlus NT-180 - Page 12

AED HeartPlus NT-180
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section I -Product Introduction
8
Configuration
Main unit
Cartridge (all-in-one type)
Cartridge (detachable type)
Electrode pads (detachable type)
Protection
Function
In case the electrode pads are not attached to the
patient, device cannot administer defibrillation shock.
If defibrillation shock is not executed by way of
pressing the flashing shock button within 20
seconds, the charged defibrillation shock energy is
automatically discharged internally.
Over-discharge protection circuit for proper
discharging of defibrillation shock energy.
Safety control circuit to prevent the malfunction of
the electric shock waveform control circuit.
Storage And
Management
Of
Data/Event
Stored content:
- Electrocardiogram.
- The number of defibrillation shocks
- Surrounding sound during the operation of the
product.
- Self-test time and result
- Time and date of product use.
Storage time:
- 5 second during analysis only, 1,000 ECGs
(analysis only)
- Surrounding sound: Max. 15 minutes
- Self-test result: Max. 3,000 cases
Data communication: USB to PC (Type B to Type A)
Data storage method:
- embedded memory

Related product manuals