2-18 Subject to Export Control, see Cover Page for details.
A. Generator Verification (cont)
18. Move the cursor to the Level dBm: field.
19. Press the F1 “Edit” Key to highlight the field. Press the ± Key, 3 Key and 0 Key to set the
field to -30.0 dBm. Press the F1 “Done” Key to store the setting.
20. Insure Power Meter Sensor range covers -30 dBm. Zero and calibrate the Power Meter.
21. Disconnect the Power Meter Sensor from the T/R Connector and connect the Power Meter
Sensor to the ANT Connector with TNC/N Adapter.
22. Move the cursor to the MHz: field.
23. Press the F1 “Edit” Key to highlight the field. Press the 2 Key to set the field to 2 MHz.
Press the F1 “Done” Key to store the setting.
24. Record Power Meter reading.
25. Repeat Steps 22 to 24 for the following frequencies:
20 MHz
100 MHz
200 MHz
300 MHz
400 MHz
500 MHz
600 MHz
700 MHz
800 MHz
900 MHz
1000 MHz
SWR CONNECTOR
26. Move the cursor to the Port: field.
27. Press the F1 “Edit” Key to highlight the field. Using the Arrow Keys ( ), select SWR.
Press the F1 “Done” Key to store the setting.
28. Move the cursor to the Level: field.
29. Press the F1 “Edit” Key to highlight the field. Press the ± Key, 5 Key to set the field to
-5.0 dBm. Press the F1 “Done” Key to store the setting.
30. Insure Power Meter Sensor range covers -5 dBm. Zero and calibrate the Power Meter.
31. Disconnect the Power Meter Sensor from the ANT Connector and connect the Power Meter
Sensor to the T/R Connector with TNC/N Adapter.
32. Move the cursor to the MHz: field.
33. Press the F1 “Edit” Key to highlight the field. Press the 2 Key to set the field to 2 MHz.
Press the F1 “Done” Key to store the setting.
34. Record Power Meter reading.
35. Repeat Steps 32 to 34 for the following frequencies:
20 MHz
100 MHz
200 MHz
300 MHz
400 MHz
500 MHz
600 MHz
700 MHz
800 MHz
900 MHz
1000 MHz