Subject to Export Control, see Cover Page for details. 2-19
A. Generator Verification (cont)
RF GENERATOR ERROR
36. Calculate the RF Generator Error by subtracting the Expected reading from the Power
Meter reading at each frequency tested on each connector.
If error is <± 3 dB, the Test Set is out of specified limits and must be calibrated.
If error is <± 1.5 dB, the Test Set should be calibrated to insure correct operation over
temperature.