D. RF Power Meter Verification (cont)
80. Record the RF Power Meter reading and Test Set Power Meter reading.
81. Set the RF Signal Generator to RF OFF.
82. Set the RF Power Amplifier to OFF.
83. Calculate the Test Set Power Meter Error by subtracting the Actual Power Meter reading
from the Test Set reading.
If error is >+0.6 dB or <–0.7 dB, the Test Set is out of specified limits and must be
recalibrated.
If error is >± 0.3 dB, the Test Set should be calibrated to insure correct operation over
temperature.