2-50 Subject to Export Control, see Cover Page for details.
F. SWR Phase Detector Calibration
TEST EQUIPMENT: Short-Open-Load VSWR Calibrator
1. With the System Menu displayed, press the F4 “Selftest” Key to display the Self Test
Menu.
2. Press the 3 Key to display the Calibration Screen.
3. Enter the Calibration Password (30203). Press the F1 “Done” Key to display the Internal
Calibration Menu.
4. Press the 0 Key (SWR PHASE DETECT CAL) to display the SWR Phase Detector
Calibration Screen.
5. Press the F4 “CAL” Key and follow instructions.
6. When the message ‘Calibration Complete’ appears, press the F5 “Enter” Key to return to
the SWR Phase Detector Calibration Screen.
7. Cycle power on the Test Set.